Measurement accuracy specified as ±0.5% of full scale usually means the stated error limit is calculated from the instrument’s full-scale reference value, regardless of the current reading. For a sensor with a 0–100 bar range, that gives an error limit of ±0.5 bar under the specified conditions. At a reading of 5 bar, the same ±0.5 bar represents 10% of the reading.
This is the practical consequence of full-scale accuracy: a percentage that looks small on a datasheet can become a large relative error near the bottom of the measurement range. To judge whether an instrument is suitable, convert its accuracy specification into engineering units, then compare that error with the tolerance required at your actual operating points.
When the manufacturer defines full scale as the measurement span, the calculation is:
Absolute error limit = accuracy percentage ÷ 100 × measurement span
For a 0–100 bar pressure sensor rated at ±0.5% FS:
Absolute error limit = 0.5 ÷ 100 × 100 bar = ±0.5 bar
The specification therefore describes an error band extending 0.5 bar in either direction. Its total width is 1 bar. It does not mean the sensor always has a 0.5 bar error, or that the error changes randomly across that entire band. The actual error may be smaller and may vary systematically with pressure, temperature, or measurement history.
To express that limit relative to a particular nonzero reading:
Equivalent percentage of reading = absolute error limit ÷ absolute reading × 100
At zero, percentage of reading is undefined because the calculation requires division by zero. The absolute error limit remains useful, which is one reason specifications based on span are convenient for instruments that measure close to zero.

For a zero-based range such as 0–100 bar, the upper range value and span are numerically identical. That makes the terminology easy to overlook. For an offset or bidirectional range, the distinction matters.
A range of −100 to +100 units has a span of 200 units, while its positive endpoint is +100 units. A specification calculated from the span produces twice the absolute error allowance of one calculated from the positive endpoint. The abbreviation “FS” alone may not establish which convention applies.
Look for the definition associated with the specification: full-scale output, full-scale range, span, rated capacity, or upper range limit. For an electrical output, “full-scale output” may refer to the output excursion corresponding to the measurement range. A 4–20 mA signal, for example, has a 16 mA output span.
The denominator is part of the specification. Two identical percentage values are comparable only when they use equivalent reference quantities and cover equivalent error components.
A percentage-of-reading specification scales its error allowance with the measured value. A pure full-scale specification gives a fixed absolute allowance for a fixed range. Consider two hypothetical instruments measuring 10 bar:
That comparison explains why percentage-of-reading performance can be attractive when measurements cover a wide operating range. However, the complete specification often includes a fixed term, such as a percentage of range, an offset, or a number of display counts.
For example, a hypothetical specification of ±(0.1% of reading + 0.02% of range), on a 100 V range at a 10 V reading, gives:
±(0.001 × 10 V + 0.0002 × 100 V) = ±0.03 V
The fixed range term becomes increasingly influential at low readings. Comparing only the percentage-of-reading term would understate the permitted error.
Specifications written as a sum normally require adding those terms as stated. They should not be combined using a statistical method merely because that gives a smaller result. Statistical combination belongs in an uncertainty analysis with justified assumptions about the contributing quantities.
In formal metrology, measurement accuracy describes closeness to the true value; it is not itself a numerical quantity. Commercial datasheets nevertheless use “accuracy” as shorthand for a stated error limit or performance specification. The useful question is which effects that number covers.
A full-scale percentage might describe nonlinearity alone, a combination of nonlinearity and hysteresis, or a broader error band that also includes repeatability and other contributions. These are different claims.
A sensor can repeat closely while consistently reading high. Likewise, a low nonlinearity figure does not establish low total measurement error if zero offset, temperature effects, or drift are excluded.
Resolution is another separate property. A display may show increments of 0.01 bar while the instrument’s stated error limit is ±0.5 bar. Those extra digits help reveal changes, but they do not establish accuracy at the last displayed digit.
A quoted accuracy often applies under reference conditions. Operation outside those conditions may introduce additional error terms. Temperature is particularly relevant because it can shift both zero and sensitivity.
Distinguish the operating temperature range from the range over which a stated accuracy or compensated performance applies. Being able to function at a temperature does not necessarily mean the instrument retains its reference-condition error limit there.
Other relevant conditions include warm-up time, supply voltage, mounting orientation, vibration, and time since calibration. Their significance depends on the instrument. Pressure measurements can also be affected by installation height and connecting lines; electrical measurements by loading and lead resistance; dimensional measurements by alignment and thermal expansion.
Dynamic behavior needs a separate assessment. Static accuracy does not establish how well an instrument captures a fast pressure pulse or follows a moving stage. Response time, bandwidth, filtering, and sampling can affect the measured result even when the static specification is adequate.
The complete measurement chain matters too. A sensor connected to a transmitter, analog input, and scaling function acquires contributions from those stages. Digital communication removes some analog conversion steps, but it does not eliminate sensor error, installation effects, or drift.
Range selection should begin with the process requirement. Identify the lowest important reading, the normal operating band, the maximum measurement input, and the error that the measurement task can tolerate.
For a purely full-scale specification, a useful preliminary relationship is:
Maximum permissible span = allowable absolute instrument error ÷ fractional full-scale specification
Suppose a hypothetical application allocates ±0.1 bar to the instrument’s specified error, and the candidate instrument is rated at ±0.5% of span. The largest span satisfying that allocation is:
0.1 bar ÷ 0.005 = 20 bar
A 0–20 bar version would therefore meet that particular allocation under the specification’s stated conditions. A 0–100 bar version with the same percentage rating would not.
This calculation is a screening tool. If ±0.1 bar is the allowance for the entire measurement system, allocating all of it to the sensor leaves no room for installation, acquisition, environmental, or other contributions.
A smaller range also has practical limits. It must cover the required measurements and accommodate the application’s pressure or input conditions. Overload survival does not imply valid measurement above the rated range. Where normal operation and occasional peaks are far apart, possible approaches include separate measurement ranges, appropriate protection, or an instrument with a suitable combined error specification.
Rescaling a display or transmitter output does not automatically change the sensor’s underlying accuracy basis. A device with a wide sensing range may retain an error contribution tied to that original range after its output is configured for a narrower interval.
For configurable transmitters, distinguish the sensor’s range limit from the configured span. The specification may use either quantity or include terms involving both. A smaller configured span can therefore make a fixed sensor error larger as a percentage of the displayed range.
Autoranging instruments introduce another variation: selecting a lower hardware range may reduce a range-dependent error term, but the applicable formula still governs. Range changes can also affect resolution, settling behavior, and overload capability.
A datasheet error limit and calibration uncertainty answer different questions. The datasheet states performance under defined conditions. Calibration reports the instrument’s response at tested points, together with the uncertainty associated with those results.
A calibration correction can compensate for a measured systematic error, provided the correction remains applicable. It does not remove uncertainty in that correction, subsequent drift, hysteresis, or effects caused by different operating conditions.
When a measurement supports a pass/fail decision, its uncertainty also affects how confidently a result can be compared with a tolerance boundary. Any acceptance rule or guard band should be defined for that task; a universal accuracy ratio cannot replace that decision.
Before approving an instrument, translate the proposed specification into a requirement that can be checked. Record the range and denominator, calculate the error at the lowest important reading and normal operating points, identify included and excluded effects, and account for the measurement chain.
A useful requirement is more specific than “accuracy better than 0.5%.” It states the allowable error in engineering units over the required measurement interval, under the relevant temperature and installation conditions. That gives suppliers a clear target and makes competing specifications meaningfully comparable.